CSI/Jewett Blood Refrigerators, Plasma Freezers, Red Cell Freezers -85C Red Cell Freezers
-85°C RED CELL FREEZERS
Jewett -85°C Super Cold red cell freezers include electronic temperature control, alarm, test and monitor functions. A built-in temperature
recorder maintains a permanent record of storage temperature. This model can be used for plasma storage as well. Optional red cell
canisters and plasma racks are available; see Accessories.

Red Cell Freezer Model LTUR13 (left), shown with Jewett
Super Cold™ ultra-low temperature chest and upright
freezer. All feature -85°C operating temperatures.
Technical Specifications
Super Cold -85°C Red Cell Freezers
Model |
Cu Ft/ Liters |
Red Cell Cap.3 |
115V, 60Hz |
Amp./Breaker*
208/230V, 60Hz |
200/230V, 50Hz |
Interior WxDxH
Inches/cm |
Exterior WxDxH
Inches/cm |
LTUR13 |
13.4 |
200 |
15.2/20.0 |
11.0/15.0 |
11.0/15.0 |
50.5" x 19.5" x 23.5"
128.3 x 48.3 x 59.7 |
79.5" x 28.75" x 33.5"
202 x 73 x 85 |
3 Standard red cell canisters. Also holds up to 240 plasma units in boxes.
*Specify voltage when ordering. Other 50Hz voltages available include 200V or 230V, AC, 50Hz.
Electronic Controller, Alarm/Monitor system
- Solid-state electronic controls with key-operated main power switch.
- Extra-large digital temperature display, resolution to 0.1°C.
- Solid-state, multiple function alarm system with self-charging battery back-up.
- Audible and visual warning of temperature deviation or power failure.
- Alarm silence button with automatic ringback function until fault is corrected.
- Independent high and low temperature alarm setpoints, adjustable in 1°C increments.
- Touchpad alarm test function physically heats probe to verify alarm readiness.
- Low battery indicator.
- Built-in voltage boost system adjusts instantly to low line current.
- Clean filter indicator light warns of dirty air filter.
- Exterior alarm contacts for connection to remote monitoring system.
- Tamper-resistant, key-operated alarm switch.
Cabinet Construction and Refrigeration
- Heavy-gauge steel cabinets with high-impact, easy to clean epoxy finish.
- CFC-free insulation, minimum 5" urethane foam to protect against high ambient temperatures.
- High strength outer door with 4" foam insulation; dual cam action hardware, lockable, with heavy-duty ball bearing hinges.
- Independent double-seal door gaskets to minimize frost build-up; individual inner doors minimize cold air loss during openings.
- Extra sound insulation to improve operating environment.
- Recessed, heavy-duty casters, built-in 1" diameter access port, standard.
- Industrial quality, hermetically-sealed compressors.
- Exclusive down-feed evaporator and automatic scrubbing cycle for maximum refrigerant flow.
- Washable condenser intake filter for unrestricted air-flow.

Solid-state controls and alarm monitor system are contained in an integrated control panel.
Touchpad controls are tamper-resistant with single multi-point key switch
for added security.
A |
Multi-position key switch for main power and alarm. |
B |
Indicator lights warn of temperature deviation or power failure. |
C |
Bright, easy-to-read digital display. |
D |
Indicator light advises when to remove and clean washable air intake filter. |
E |
Indicator light advises when to replace self-charging alarm system battery. |
F |
Alarm test button initiates test function to physically heat probe; no door opening is required. |
G |
Alarm silence button quiets audible alarm; alarm ringback continues until temperature returns to
normal. |
H |
Touch pad setpoint controls for operating temperature, high alarm and low alarm limits. |
I |
Indicator light warns of low voltage, and indicates built-in voltage boost system is activated. |